Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Table of Contents

Safety Precautions:

  • You must complete required lab safety training before starting this procedure.
  • You must complete required animal use training and listed on IACUC protocol before starting this procedure.
  • If this is your first time doing this procedure, ask to be trained by an experienced lab member.  If you have not done this in a while, you should ask for a refresher.
  • Also review the following material:
    • Kuwajima M, Mendenhall JM, Harris KM (2013) Large-Volume Reconstruction of Brain Tissue from High-Resolution Serial Section Images Acquired by SEM-Based Scanning Transmission Electron Microscopy. Methods Mol Biol (Nanoimaging: Methods and Protocols) 950:253-73. (PDF)
  • Further reading:
    • Hayat MA (1981) Fixation for electron microscopy. Academic Press
    • Karnovsky MJ (1965) A formaldehyde-glutaraldehyde fixative of high osmolality for use in electron microscopy. J Cell Biol 27:137A-138A. (PDF)
    • Tao-Cheng JH et al. (2007) Structural changes at synapses after delayed perfusion fixation in different regions of the mouse brain. J Comp Neurol 501:731-740. (PDF)
    • Dehghani A et al. (2018) Nuclear expansion and pore opening are instant signs of neuronal hypoxia and can identify poorly fixed brains. Sci Rep 8:14770. (PDF)
  • Before starting, even if you have done this procedure before,
    • read this protocol entirely
    • If you are using adult rats, review the instruction movies on the tracheotomy procedure.
    • review relevant Safety Data Sheets and Harris Lab SOP (also see below)
    • ensure you have all reagents and supplies listed belowensure all equipment is in good working order
      • Isoflurane vaporizer must be serviced annually; check tubing, surgical instruments, etc.
    • have all waste containers ready (also see Clean-up)
    • plan your schedule well so that you wouldn’t have to rush
  • Review SDS and Harris Lab SOP for the following hazardous chemicals used in this procedure:
    • Ethanol: flammable; irritant (eye)
    • Formaldehyde
    • Glutaraldehyde
    • Hydrochloric acid
    • Isoflurane
    • Sodium cacodylate: carcinogen; irritant (skin, eye); skin permeator
    • Sodium hydroxide
  • Other hazards associated with this procedure include:
    • Physical: compressed gas cylinder (O2/CO2)
    • Sharps: Surgical scissors, scalpel, needles
    • Biological: Live rats/mice, animal waste and soiled cage bedding
  • The following Personal Protective Equipment is required for this procedure:
    • Lab coat
    • Nitrile gloves (double-layer required; regularly check for holes)
    • Eye goggles
    • Mask
    • Face shield
  • Place a piece of absorbent sheet on the work surface before starting the procedure.  When done, discard into the “Solid Waste – No UA” bag

 

Reagents, Supplies, Equipment

Personnel Protective Equipment:

  • Eye goggles, face shield
  • Gloves (nitrile)
  • Lab coat
  • Mask
  • Plastic Apron (optional)

Air Pressure System:

  • Air pump: Perfusion Two (MyNeuroLab.com)
  • 1-L glass bottle (Kimax, or equivalent) × 1
  • #6 two-hole stopper × 1
  • Sphygmomanometer with a hand pump (cut off the cuff)
  • Air filter (Parker Hannifin 9933-05-AAQ)
  • Assorted tubing, connectors, and clamps

Oxygenation System:

  • O2/CO2 (95%/5%) cylinder with regulator and support
  • Assorted tubing, connectors, and clamps

Anesthesia System:

  • Isoflurane (Animal Health International 19632158; stored at RT in well-ventilated area)
  • Large glass desiccator, wad of 4-5 Kimwipes, Pasteur pipet
  • Matrx VIP3000 Isoflurane Vaporizer (Stoelting)
  • A small animal ventilator (Harvard Apparatus Model 683, or equivalent)
  • Nose cone (Kent Scientific)
  • endotracheal tube (a modified 16-gauge hypodermic needle) with Y-connector
  • Assorted tubing, connectors, and clamps
  • Optional: 10”×4”×4” Plexiglas anesthesia box (Stoelting)

Perfusate System:

  • Deep water bath (e.g., VWR Digital Unstirred Water Bath, L×W×D = 127/8"×1113/16"×529/32"; VWR 89032-216)
  • Lead donuts × 2
  • #6 stoppers: three-hole × 1, two hole × 1
  • 3-way valve × 1
  • Flow regulator × 1
  • 13-gauge needle with 60° bevel with cork disk ½ inch from tip (to restrict depth of penetration; use a16-gauge for juvenile rats)
  • Air stone × 1
  • Assorted tubing, connectors, and clamps

Waste Collection System:

  • 4-L vacuum flask (preferably plastic-coated) × 1
  • #11½ one-hole stopper × 1 
  • Vacuum line filter (Whatman VACU-GUARD, VWR 28137-858), 1 per day of perfusions
  • Assorted tubing and connectors

Dissection/surgical instruments:

  • Stainless steel dissection tray with Styrofoam board cut to fit snugly lengthwise in the tray with ~1-inch space on one side. 
  • Assorted dissecting scissors
  • Silk suture (or cotton tread from fabric store can be used instead)
  • Gauze pads
  • Forceps
  • Hemostats
  • Scalpel and blades
  • Knife (to decapitate rat)
  • Bone Rongers
  • Spatulas
  • Applicator sticks
  • Pins to secure animal to Styrofoam board


For Preparation of Perfusates:

  • Calcium chloride dihydrate (CaCl2·2H2O; Sigma-Aldrich 223506; stored at RT)
  • d-Glucose (Sigma-Aldrich G7528; stored at RT)
  • Formaldehyde (20% aqueous solution; Ladd Research 20304; 100 ml bottles; stored at RT)
  • Glutaraldehyde (50% aquieous solution; Ladd Research 20211; 100 ml bottles; stored at 4°C)
  • Hydrochloric acid (HCl; 1N aqueous solution; for adjusting pH)
  • Magnesium sulfate heptahydrate (MgSO4·7H2O; Sigma-Aldrich M5921; stored at RT)
  • Potassium chloride (KCl; Sigma-Aldrich P9333; stored at RT)
  • Sodium bicarbonate (NaHCO3; Sigma-Aldrich S6297; stored at RT)
  • Sodium cacodylate trihydrate (Ladd Research 20305; stored at RT)
  • Sodium carbonate (Na2CO3; Sigma-Aldrich 223530; stored at RT)
  • Sodium chloride (NaCl; Sigma-Aldrich S7653; stored at RT)
  • Sodium hydroxide (NaOH; 1N aqueous solution; for adjusting pH)
  • Water (double-distilled, ASTM type I, WFI [water for injection], or equivalent; e.g., Fisher 91-502-5)
  • 2- or 4-L glass beaker × 2
  • Magnetic stirrer × 2
  • 100-ml glass beaker and a disposable pipet (for dissolving sodium carbonate [Na2CO3])
  • 1- or 2-L graduated cylinder (depending on the final volume)
  • pH meter (and calibration standards)
  • 1- or 2-L glass bottle (Kimax, or equivalent) × 2
  • Vacuum filtration system (pore size = 0.22 µm; e.g., VWR 10040-468 – this one screws onto the Kimax glass bottles)
  • WESCOR VAPRO 5520 Osmometer (with calibration standards, filter paper disks, pipet, and pipet tips)

Worksheets for Record Keeping:


Perfusion Apparatus Assembly

The perfusion apparatus consists of five sub-systems (Fig. 2): 1) air pressure (Green), 2) perfusate (Blue), 3) oxygenation (Red), 4) anesthesia (Pink), and 5) waste collection (Brown).

The air pressure system (Figs. 2 and 3) consists of a modified Perfusion Two air pump, air tank, and a modified sphygmomanometer.  The air tank (1-L Kimax bottle; 4 in Fig. 3) is sealed with a #6 two-hole stopper.  A Y- (or T-) connector is attached to one of the two holes a hole in the stopper, to which two pieces of tubing are attached to connect the air tank to #2 and #3 ports of the Perfusion Two air pump (2 and 3 in Fig. 3, respectively).  A hand pump (the bulb detached from the cuff of a sphygmomanometer; 5 in Fig. 3) An in-line air filter is connected between the #3 port and the air tank.  A  is attached to the other hole in the stopper.  #1 port of the air pump (1 in Fig. 3) is connected to a hand pump (the bulb detached from the cuff of a sphygmomanometer; 5 in Fig. 3), the sphygmomanometer (7 in Fig. 4), and perfusate bottles (see below). 

The perfusate system (Figs. 2, 4, and 5) consists of a water bath, two 2-L bottles (one for KRC and the other for the fixative), perfusion needle, assortment of  tubing, connectors, clamps, and valves.  Lead donuts (6 in Fig. 4) are placed around the necks of the bottles to keep them from floating and tipping over in the water bath as the fluid in them is exhausted.  The fixative and the KRC bottles are sealed with #6 stoppers (Figs. 5 [two-hole stopper for fixative bottle] and 6 [three-hole stopper for KRC bottle]).  The holes are used to connect the bottles with 1) air pressure system, 2) oxygenation system (KRC only), and 3) a 3-way switch and tubing leading to the perfusion needle.  KRC and fixative are driven during perfusion by the air pressure system (see above).  It is very important that the stoppers and the inside of the neck of the bottles be completely dry before inserting the stoppers to prevent the stoppers from blowing out under pressure.  When the apparatus is assembled prior to each perfusion, it should be tested for the pressure of up to 200 mmHg to ensure the connections throughout the system will withstand pressure and not blow apart.  Superglue is used to secure joints between different sized tubing.  

Connecting the perfusate bottles to the air pressure system:

Attach the long end of Y-connector into one of the three holes in the stoppers (2 in Figs. 5 and 6).  Use a short piece of Tygon tubing on one side of Y-connector (2b in Figs. 5 and 6) to connect the two perfusate bottles.   On one of the stoppers, another piece of tubing (2a in Fig. 5) is attached to the unused side of the Y-connector to connect the sphygmomanometer (indicator; 7 in Fig. 4) and the #1 port of Perfusion Two air pump (8 in Fig. 4 connects to 1 in Fig. 3) via another Y-connector.  On the other stopper, a short piece of tubing is attached to the unused side of the Y-connector for use as a pressure relief valve while KRC and fixative are gassed with O2/CO2 (2a in Fig. 6 and 5 in Fig. 4).  This pressure relief tubing is folded over and clamped with a tubing clamp when pressure is applied by the air pump during the perfusion.  

Connecting the perfusate bottles to the perfusion needle:

Through the second hole of the bottle stoppers, run a piece of Tygon tubing (4 in Figs. 5 and 6; must be long enough to reach the bottom of the bottles), which is then connected to the three-way valve (3 in Fig. 4).  The third connector on the three-way valve is connected via a luer-lock connector to IV tubing, to which the perfusion needle (4 in Fig. 4) is attached.  A flow regulator/clamp is installed on this segment of the tubing.  The length of the IV tubing from the bottles to valve and then the needle should be kept as short as practical: the longer the distance between the water bath and the animal the greater the drop in temperature of the perfusates.  (The temperature of the water bath may need to be adjusted in order to deliver fixative at 37°C, depending on the length of tubing, ambient air temperature, etc.)  

Connecting the perfusate bottles to the oxygenation system:

Through third hole in the stoppers, run a piece of Tygon tubing (3 in Figs. 5 and 6; long enough to reach the bottom of the bottles).  Attach air stones (2 in Fig. 4) on the bottle side of the tubing.  The other side of tubing is connected to O2/CO2 cylinder-regulator, with a clamp in between (1 in Fig. 4; clamp not shown).

Connecting the oxygenation system to the anesthesia system:

The anesthesia system (Figs. 2, 7 and 8) consists of an isoflurane vaporizer driven by an O2cylinder (or the O2/CO2 cylinder from above), a nose cone, endotracheal tube, and a ventilator to ventilate the anesthetized animal during the dissection and the initial stages of perfusion.  The vaporizer and ventilator will not be used for mice or juvenile rats.  A large glass desiccator is used to anesthetize the animal prior to putting on the nose cone (an optional small Plexiglas chamber connected to the vaporizer can be used for this purpose).  For juvenile rats and mice, a 15-ml conical tube with gauze pads containing ~1 ml isoflurane is used as the nose cone.

Connecting the oxygenation system to the anesthesia system:

Connect O2/CO2regulator to the O2 input on the back panel of vaporizer (behind the O2/CO2 pressure flow control; 6 in Fig. 7).  Connect the isoflurane output (8 in Fig. 7) to a Y-connector.  One side of the Y-connector is attached to a piece of silicone rubber tubing (9 in Fig. 7) leading to the nose cone (4 in Fig. 8).  Use another piece of silicone rubber tubing to connect the other side of the Y-connector to air intake of the ventilator (11 in Fig. 7).  Output from the ventilator (12 in Fig. 7) is connected to the endotracheal tube (1 in Fig. 8; a blunted 16-gauge needle) via a Y-connector.  Exhaust line from the endotracheal tube (2 in Fig. 8) connects back to the exhaust port of the ventilator (13 in Fig. 7).  Make sure to label the exhaust line.

The waste collection system (Figs. 2 and 9) is used to remove blood and perfusate from the dissection tray as it accumulates, and consists of a 4-L vacuum flask attached to vacuum line.  A #11½ one-hole stopper is used to seal the flask (Fig. 9).  Through the hole, attach a short segment of a plastic serological pipet (long enough to pass the vacuum port of the flask), to which Tygon tubing is connected (1 in Fig. 9).  Place the other end of this tubing to the dissection tray (5 in Fig. 8).  The flask is connected to vacuum line via heavy rubber vacuum tube (3 in Fig. 9).  An in-line filter (2 in Fig. 9) is placed to prevent waste liquid from getting into vacuum line.  Replace the in-line filter after each day of perfusions, or when the filter gets wet.  After the procedure the contents of the vacuum flask are transferred to appropriate disposal containers in accordance with local institutional policies.


Preparation of Perfusate Solutions

Prefix perfusate (Krebs-Ringer Carbicarb buffer, or KRC)

Reagent

[final] mM

F.W.

For 0.5 L

For 1 L

For 2 L

Purified water to start with

-

-

~350 ml

~800 ml

~1600 ml

NaCl

118.0

58.44

3.448 g

6.896 g

13.792 g

KCl

4.7

74.55

0.175 g

0.350 g

0.701 g

CaCl2·2H2O

2.0

147.02

0.147 g

0.294 g

0.588 g

MgSO4·7H2O

4.0

246.48

0.493 g

0.986 g

1.972 g

NaHCO3

12.5

84.01

0.525 g

1.050 g

2.100 g

D-glucose

11.0

180.16

0.991 g

1.982 g

3.964 g

Na2CO3*

12.5

106.00

0.663 g

1.325 g

2.650 g

Oxygenated "Krebs-Ringer Carbicarb (KRC)" buffer is used to flush blood cells prior to fixative perfusion.  It is prepared from dry reagents as described here on the day before the procedure and stored at RT overnight.  Make a fresh batch before each day of perfusions.  One-liter of KRC should be sufficient for up to eight mice or juvenile rats, or 2 L for up to four adult rats.

*IMPORTANT: DO NOT directly add Na2CO3 (sodium carbonate) as a dry solid reagent!!  Dissolve Na2CO3 in ~50 ml of water separately from the other reagents.  When the other reagents have dissolved, slowly (1-2 ml at a time) add the Na2CO3 solution with continuous mixing.  As Na2CO3 is added, it will be necessary to adjust the pH between 7.5 and 8 with 1N HCl to prevent precipitation.  After all of the Na2CO3 has been added, allow the solution to mix for several minutes before adjusting the pH to 7.35-7.40, then bring to the final volume with purified water.  Filter through a vacuum filtration system and place in a 1- or 2-L bottle.  

Measure the osmolality (see 3.3.below), and if necessary adjust it to 300-330 mmol/kg.  If you were careful in weighing the reagent, the solution should be within this range and no adjustment will be needed.  KRC is warmed to 41°C in a water bath (37°C at the tip of perfusion needle) and gassed with O2/CO2 (95%/5%) for at least 30 min before use.  

Fixative

Reagent

[final]

F.W./Stock

For 0.5 L

For 1 L

For 2 L

Purified water to start with

-

-

~300 ml

~600 ml

~1200 ml

Na cacodylate*

100 mM

214.0

10.700 g

21.400 g

42.800 g

CaCl2·2H2O

2 mM

147.02

0.147 g

0.294 g

0.588 g

MgSO4·7H2O

4 mM

246.48

0.493 g

0.986 g

1.972 g

formaldehyde

2.0%

20.0%

50 ml

100 ml

200 ml

glutaraldehyde**

2.5%

50.0%

25 ml

50 ml

100 ml

Our standard perfusion fixative, 2% formaldehyde + 2.5% glutaraldehyde, is shown here.  Each adult rat will require about 2 L of fixative, while at least 500 ml will be used for each mouse or juvenile rat.

*IMPORTANT: Sodium cacodylate is a known carcinogen.  Open the bottle and weigh only in a chemical fume hood.

**IMPORTANT:  Three days prior to fixative preparation, remove 50% glutaraldehyde stock from refrigerator and leave at RT.  

The stock buffer minus the aldehydes is prepared the day before use and stored at RT overnight.  After the salts are dissolved, adjust pH to 7.35, and store at RT.  On the day of use, move the solution to a fume hood before adding the aldehydes and bring to the final volume, filter through a vacuum filtration system and place in a 1- or 2-L bottle.  Measure the osmolality (see 3.3.below), which should be 900-1100 mmol/kg.  Warm the solution to 41°C (37°C at the tip of perfusion needle) in water bath before use.  

Measuring osmolality with WESCOR VAPRO 5520 osmometer

See here.


Anesthesia

The objective is to anesthetize the animal so that it does not experience pain during the dissection but is still alive with the heart beating at the time of perfusion.  Weigh the animal and record pertinent information on the Perfusion Worksheet.  A large glass desiccator jar with a wad of 4-5 small Kimwipes is used as an anesthesia chamber for adult rats.  A glass vial with perforated lid (containing a piece of Kimwipe) in a plexiglass chamber can be used for juvenile rats or mice.  Five minutes prior to starting, add 1.5 ml of 100% isoflurane to the Kimwipes below the perforated plate and put the top on the desiccator.  Place the animal in the desiccator after the atmosphere has equilibrated for 5 minutes.  For adult rats, open the O2/CO2 regulator, set the vaporizer at 5% isoflurane (press the release button [5 in Fig. 7] and turn the dial control [4 in Fig. 7]), and adjust the O2 flow to 400 ccm (6 [control] and 7 [indicator] in Fig. 7).  Also turn the ventilator on, and adjust the stroke rate (15 in Fig. 7) to 120 breaths/min and the tidal volume (14 in Fig. 7) to 1.5 cc.  Within about 1.5 to 2 minutes inside the desiccator, the animal should begin to stumble and fall over (0.5-1 min for mice or juvenile rats).  Anesthesia should have reached the stage where the animal is non-responsive to toe pinch in about 2.5-3 minutes (1-2 min for juveniles).  At this point, place the animal on the dissection tray with its nose inside the nose cone.  Secure the animal to the Styrofoam board with T-pins.  

If the Plexiglas rodent anesthesia chamber is used instead of the desiccator, place the animal in the chamber attached to the isoflurane vaporizer.  Set the vaporizer at 5% isoflurane and adjust the O2 flow to 400 ccm.  The rat should become anesthetized in about 4 minutes.  Check this by rotating the chamber slightly and observing if the animal tries to adjust its stance, i.e., tries to stay on its feet.  The vaporizer should be connected separately to the box, nose cone and ventilator using Y-connectors and tubing.  If the nose cone is connected to the box via the box’s exhaust as shown in some vendor’s diagrams, the animal will not continuously receive the same level of isoflurane via the nose cone after the box is opened.  

For perfusing mice or juvenile rats (or if a vaporizer and a ventilator are not available), add 0.5 ml of isoflurane to gauze pads in the bottom of a 15-ml conical tube and use as a nose cone.  Make sure that the anesthetic would not drip out of the tube.  In this case, tracheotomy will not be performed, and therefore the time between cutting the diaphragm and perfusion with the fixative must be as short as possible (30 sec or less) to minimize hypoxic damages to the brain tissue. 

Division of Labor during Tracheotomy and Perfusion           

This procedure is best performed by a "surgeon" and "assistant."  The assistant has a crucial role in preparing the perfusion apparatus while the surgeon is anesthetizing the animal and performing the tracheotomy, and in regulating the perfusion pressure and changing from KRC to fixative in the early stages of the perfusion.  After the heart is sufficiently fixed to hold the needle in place, the rest of the perfusion could be performed alone by either person.  In the Step-By-Step procedure outline below, the minimal role of the assistant is shown in italics.

 Before Before the animal is anesthetized, the perfusion apparatus must be completely pressure-tested, the perfusion solutions brought to 41°C (such that they are at 37°C at the tip of the perfusion needle), and KRC gassed with O2/CO2 (95%/5%) for at least 30 min.  The assistant must also flush fixative to remove air bubbles from the fixative line to the point just past the 3-way switch, and then KRC throughout the system to flush out the fixative and any remaining bubbles.  Then the assistant must get the pressure stabilized at 80 mmHg and be ready to begin the flow of KRC to allow about 10 ml (enough to clear the tubing and begin perfusing with warm KRC) before the surgeon inserts the perfusion needle into the heart.  The assistant must be ready to switch from KRC to fixative within 5 sec after the left ventricle is penetrated.  For mice and juvenile rats, this switch must occur immediately after the needle insertion.  The assistant will change the pressure as required to ensure good fixation at least as long as the surgeon is tending to the animal.

Step-By-Step Procedure:

All procedures are performed in a well-ventilated fume hood or on a necropsy table with down draft.  Appropriate protective clothing, including gloves and eye protection, are worn.  All waste is contained and disposed of according to local regulations.

The minimal role of the assistant is shown in italics.

Steps to be skipped for mice or juvenile rats are indicated by A (for "adult rats only").  They do not require tracheotomy and artificial ventilation.  Therefore, time between cutting the diaphragm (i.e., loss of breathing) and perfusion with fixative must be as short as possible (30 sec or less) to minimize hypoxic damage to the brain tissue.

  1. Place a piece of absorbent pad under the dissection tray.
  2. Lay out surgical tools for perfusion.
  3. After filling the perfusate bottles, close them with the rubber stoppers.  Open the pressure release clamp, and close three-way valve and flow regulator on the perfusate line.
  4. Connect Tygon tubing from the perfusate bottles labeled "#1" to the #1 port on Perfusion Two pump.
  5. On Perfusion Two pump, set Air Tank switch at "Disconnect" and Perfusion Pressure switch at "Hold" positions.  Adjust air pressure dial to about 80 mmHg.
  6. Warm the perfusate in water bath to 41°C, and oxygenize it for 30 min.
  7. Close pressure release clamp on perfusate bottles.
  8. Before the animal is anesthetized:
    1. Make sure to have a new in-line filer for the vacuum line of the waste collection system.  Turn on the vacuum. Attach a waste suction line to the dissection tray with a piece of tape.  Check the system by rinsing the tray with some water.
    2. The O2 /CO2 supply to the perfusate bottles must be closed and tightly clamped.
    3. The pressure release tubing on perfusate bottles must be clamped tightly.
    4. Turn on Perfusion Two pump (the main switch is on the back panel) with Air Tank switch at "Disconnect" and Perfusion Pressure at "Hold" positions.  Compressor will start pumping air until the pressure reaches about 80 mmHg in air tank (This will not be indicated on sphygmomanometer, which measures the pressure in perfusate bottles). Turn Air Tank switch to "Connect" and Perfusion Pressure switch to "Release" positions.  Then increase the pressure to 200 mmHg to test the perfusate system (Always check with the sphygmomanometer - the indicator on the pressure control dial is approximate).  Adjust air pressure to 80 mmHg by first turning down the pressure control dial, and then by opening the valve on hand pump.
    5. Open the three-way valve and flow regulator to flush the perfusate system, first with fixative and then the KRC, to remove all air bubbles from the tubing.  Bubbles tend to get stuck at the three-way valve, connectors, and flow regulator.  The waste fluid is drained into the dissection tray, and taken up by the waste collection line.  Once this is done, close the flow regulator, with three-way valve open to KRC.
    6. For mice or juvenile rats, prepare a nose cone by placing several plies of gauze into a 15-ml conical tube.  Add a small amount (~1 ml) of isoflurane to gauze and recap the tube until use. 
  9. Anesthetize the animal in a large desiccator jar with a wad of 4-5 small Kimwipes.  Five minutes prior to starting, add 1.5 ml of isoflurane to the Kimwipes below the perforated plate and put the lid on the desiccator.  Make sure the animal does not come in contact with liquid isoflurane.
  10. (A) Start the vaporizer set at 5% isoflurane  and the O2 flow at 400 ccm, and the ventilator running at 120 breaths/min and a tidal volume of 1.5 cc.
  11. Remove the animal from desiccator and test the animal’s anesthesia level with a toe pinch.  A sufficiently anesthetized animal will not respond.
  12. Place the animal on the dissection tray and attach its nose to the nose cone.
  13. Secure the animal to Styrofoam board with T-pins.
  14. (A) Make a chin-to-sternum incision.
  15. (A) Isolate the trachea, elevate the trachea by placing the tip of a pair of curved hemostats under it to place a piece of silk suture (or cotton thread) around it proximal to the larynx and tie a loose knot in it.
  16. (A) Make a lateral cut across trachea just below larynx and above the suture, being careful not to cut completely through trachea or to let any fluids get in it.
  17. (A) Reduce the O2 flow to the vaporizer 100 ccm and reduce the isoflurane  level to 4%.
  18. (A) Remove the nose cone.
  19. (A) Immediately insert the endotracheal tube into the trachea and tie the suture.  Secure the tube by taping it to a rolled up paper towel just above the animal’s head.
  20. (A) Clamp the exhaust line of the endotracheal tube or place your right index finger over the exhaust port on the ventilator for 2-3 breaths to create an artificial "sigh."
  21. Open the chest by making a small hole below sternum and carefully cut laterally just under the diaphragm.  Clamp the sternum with a pair of hemostats.  Then carefully cut the diaphragm along the ribcage.  Cut up the two sides of the chest, keeping the scissors away from the heart and lungs, until the incisions reaches just below the forelimbs.  Lift the sternum up and turn over the neck to expose the chest cavity.  Clean away the pleura to release the heart and lungs.
  22. (A) Repeat the artificial sigh sequence.
  23. (A) With artificial respiration complete (heart beating and lungs moving in and out), confirm that the assistant has the perfusion system ready to go.
  24. Make an incision with iris scissors in the right ventricle.  At this point, the assistant opens the flow regulator for KRC.
  25. Take hold the apex of the heart with your thumb and forefinger or a good pair of tissue forceps.  With the warm KRC flowing at 80 mmHg, insert the perfusion needle into left ventricle until it stops at the cork disk and hold it there.  KRC will flush the blood through the systemic circulation and out the right ventricle.  There should be no fluid entering the endotracheal tube and ventilator tubing.
  26. In 3-5 seconds, switch the 3-way valve to begin the flow of fixative (For mice or juvenile rats, do this immediately after the needle is inserted).  Increase the pressure to 120 mmHg. For mice or juvenile rats, take the nose cone off at this point.
  27. (A) After 15 seconds gradually increase the pressure to 180 mmHg and maintain this pressure for 5 min.  During this time the animal should be having spasms and contractions and should start to stiffen.  
  28. Clip the chin and observe the fixative oozing from the cut.  Other signs of a good perfusion include: (1) feet and tail are pale white, and (2) the tail is stiff down to its tip.
  29. When the heart is fixed, usually within a couple of minutes, and the needle is secure the surgeon can let go of it, or if need be, prop it up with T-pins so that it won’t move.
  30. (A) Turn off the ventilator and vaporizer, and remove the endotracheal tube.
  31. After 5 min of perfusion at a higher pressure (180 mmHg for adult rats or 120 mmHg for mice), decrease the perfusion pressure to 80 mmHg by opening the valve on hand pump and then adjusting the pressure control dial.  Maintain this pressure for about 50 min (or 20 min for mice or juvenile rats).  At the end of this period, approximately 1800 ml (or 500 ml for mice or juvenile rats) of fixative should have been used.  DO NOT ALLOW THE FIXATIVE BOTTLE TO RUN OUT OF FIXATIVE BECAUSE THAT COULD CAUSE AIR TO BE PERFUSED INTO THE BRAIN.  Reduce the flow of fixative even to the point of cutting it off, if necessary, but the duration of perfusion with the fixative should be at least 60 min (or 30 min for mice or juvenile rats).  Note that the minimum pressure adjustable by Perfusion Two is about 70 mmHg.  If it must be reduced further, turn off Perfusion Two (with the Air Tank switch at "Connect" and Perfusion Pressure switch at "Release") and use the hand pump.  Make sure the fixatives still drips out of the clipped chin after reducing pressure. 
  32. Disconnect the perfusion needle and remove the animal from the Styrofoam board.  Leave the animal under fume hood for at least 1 hr before dissecting the brain out.
  33. Turn off Perfusion Two pump, release remaining pressure through vent on perfusate bottles.  Use the remaining KRC to rinse the Styrofoam board and tray.  Save the remaining fixatives to store vibraslices later.  Rinse the perfusate bottles with diH2O, then flush the perfusate line with diH2O.  Collect all waste liquid into the waste flask.
  34. Remove the head by cutting through the neck with a sharp knife and dissect out the brain from the skull being careful not to nick it with tools.  Place the brain in a container filled with the fixative for overnight at RT before it is examined and vibrasliced.
  35. Transfer the brain into a petri dish filled with 0.1M sodium cacodylate buffer, and examine the brain under a dissecting microscope to determine whether or not small red veins are visible.  A well-perfused brain will appear light gold and have no red veins whereas a poorly perfused will have a pink tinge.  Take images for record.  Transfer the brain back into the fixative.
  36. If the surface of the brain appears to be well perfused, slice it with a vibrating blade microtome.  For our standard EM processing, it should be sliced in phosphate buffer (0.1 M, pH = 7.4) at 70-µm thickness in the parasagittal plane (can be up to 100 µm thickness).  Collect vibraslices in 24-well plates containing 0.1 M sodium cacodylate buffer.  Dissect the area(s) of interest with a fine dissection knife in 0.1 M sodium cacodylate buffer and embed into 7% agarose for further processing.  If not immediately processed for EM, store the vibraslices in the same fixative at RT in 24-well plates (sealed with Parafilm). 

Clean-up

  • Perfusate bottles:
    • Remove the tubing from the bottles.
    • Empty excess KRC into sink, and add ~100 ml RO water.
    • Empty excess fixative into a plastic 5-gallon waste container, rinse once with some RO water (collect waste into the same container).  Then add ~100 ml RO water.
    • Re-attach the tubing and run RO water to clean the tubing. Collect waste into the 5- gallon container.
    • Detach the tubing and hang dry.
  • Surgical instruments:
    • Surgical tray and the styroform board should be rinsed with water (use the waste collection system), then disinfected with bleach
    • Surgical instruments should be soaked and washed in cold running water (worm water will clog blood), then disinfected with 70% ethanol.  Air-dry completely before storage.
  • Solid waste:
    • PPE (gloves, mask, face shield, etc.) à Place in the solid waste (no UA) bag.
    • Lab supplies (absorbent pad, Kimwipes, tubes, etc.) à Place in the solid waste (no UA) bag.
    • Carcasses à Double-bag and place into the carcasses freezer in the NHB vivarium.
    • Animal cage à Return to the NHB vivarium.
  • Liquid waste:
    • Collect in plastic 5-gallon container.  Request for pick-up by EHS.