This is a summary of normal IGRINS use for queue observers at Gemini South Telescope.

Observing timelines: 

Start of Night:

1. Wake up the observing iMac and login.

2. From the iMac: Screen share to the ICC. From the screen sharing program, open another screen share connection to the DSC. If this one isn't loading well, change the "view" tab to "full quality". After it loads, change back to "adaptive". Everything else following will be done on the ICC or DSC through the screen sharing.

3. On the DSCTurn on the IG proxy.

> cd ~/igrins_admin
> sh run_tcsproxy_gemini.sh
When it launches, the proxy window will pop up. Leave the rotator box checked.

If the VII connection is working, you should see the heart symbol beating and correct coordinates displayed. If it is red or otherwise, you have to ask the SOS to restart the VII.

4. On the ICC: Close the HPK and IGRINS Hub windows on the Desktop if they happen to be open. There are several scripts you must begin now, each in a new terminal window.

 a. Mount the detector computers

 > cd ~/igrins_admin
 > sh mntrun

b. Upload the temperatures to the web with firebase

 > cd ~/igrins_admin
 > sh run_hk2firebase.sh

c. Turn on the IG software (HKP house keeping package, SCV slitview camera, DTP data taking package) by opening a new terminal and typing:

> cd ~/igrins_admin
> sh run_icshub.sh
When it launches, the "IGRINS Hub" window will pop up.

You should see the "check Mount" and "check Proxy" change from "..." to "Okay" status.

Start the HKP (click "run HKP"). In the HKP gui, click the “periodic monitoring”  and, below that, the “Send Alert (T>20)” buttons.

Once you see that the HKP status says "connected" in the Hub, start the SCP (click "run SCP"). You should see the SCP window pop open.

Once you see that the SCP status says "connected" in the Hub, start the DTP (click "run DTP"). You should see the DTP window pop open.

d. Turn on the ICC audio  

> cd ~/igrins_admin/misc
> sh run_ics_monitor.sh
Once you run this, then you can open a browser on the observing iMac. Make sure to push play on the bottom and that the speaker on the iMac is on!

e. Fill the digital logs

> cd ~/igrins_admin/misc
> sh run_log_uploader.sh 2018A_Gemini

 It will pop open a web browser with the log opened.

5. On the ICCCheck the software

In the SCV, check if you are receiving the TCS information or not (blue values on the upper right, including RA, Dec). 

Note -  TCS will not be running without your TO. Images and cals can still be taken but headers will be incomplete.

In the Status box, the following lines (for the SVC, and then for both H and K in the DTP) should appear:

6. Test Exposures

Change the observer name to your Last Name. No single quotes or special characters. Commas are ok.

Change the object name to 'test'.

Take a test image on both SVC and with the HK detectors to test readout.

If you see noise pattern or low-resolution image, repeat the short exposures or contact observing support.

Calibrations:

The recommended calibrations are listed in the following table, and the procedure below. It should take ~25 minutes in total.

You should get 10,000 counts in the H band and 8,000 in the K band. You must have more than 4,000 and no more than 20,000.

 

Typeexposure time (s)numberNumber of Fowler Reads
Dark10106
Flat off10106

Flat on

10106


※ How to put the program ID into fits headers of IGRINS daily/SKY calibration frames

'Daily calibration' is usually the first frames of each nights.

Procedure to put Program ID:

1. Darks

Check all the lights are off around the telescope.

Setup:

Using the DTP, take darks:

 Don't forget to start the paper observing log.

2. Flats

Ask your SOS to:

A. FLAT OFF: Take flat offs before the flat lamps are on (required for pipeline). SOS are familiar with term "Lamp Off"

Wait for your SOS to make sure the telescope is ready.

Using the DTP:

If you see emission lines in the spectra, then a light has been left on somewhere.

B. FLAT ON: Take flats with the flat lamps on (also required for pipeline).

Kindly ask your SOS to turn on the GCAL lamps on (the ND1 filter should still be in place, and the fold mirror directing the light to the visitor IGRINS port). Quartz lamp.

Using the DTP:

How to use IGRINS at the Gemini South Telescope:

1. Telescope Pointing Procedure

A. Ask your SOS to slew to the target (give program ID and target ID).

B. Once on target and the SOS has finished setting up the guiding, take a test image with the SVC (see below).

C. Check that you are pointing to the correct field.

2. Using the SVC

You will break guiding if you move more than 30"

A. Taking an SVC image using the SCV:

B. Using the SCV to move the telescope:

C. Using the SCV to move your target to the reference position:

For further information about the SCV options (such as scaling, Gaussian fitting, etc.) see the "How to Use IGRINS software" page.

3. Using DTP to obtain spectra

A. Move your Target onto the slit

Either use the SCV to nudge your target onto the slit, or begin the process described in the C section below to use the software to put it in the A or B position.

B. Taking H/K spectra:

C. Nodding or using AB positions:

Nodding will produce better results in data reduction. Using an ABBA pattern is the most common. 

Nightly Observing Procedures:

1. Focus

The focus does not need to be changed at GST.

2. Obtaining Spectra for Targets and Standards

The standard procedure for obtaining data with IGRINS should be:

The most common observation will be to obtain an ABBA (or quad) orientation. This means the object is placed on the slit in the A position for the first spectra, then moved to B for a second (then again in reverse). The resulting 4 spectra using this ABBA combination will provide optimal background subtraction for point sources. There are simple scripts available for performing an ABBA observation (see the Additional Options section). Another common observation, particularly for extended sources, will be On-Off. To perform On-Off observations, obtain a spectra with your target on the slit, then another spectra with your target moved entirely off the slit.

3. Obtain Sky Spectra

You need at least one 300s sky frame each night for the PLP to work correctly. A faint target can be substituted, but if you only observe bright objects then get a sky.

'SKY' is usually the last object of each nights. Telescope setting for SKY requires 5-10 min to be prepared, so we recommend to take SKY in the end of the night slightly after twilight.

SKY Procedure:

Additional Options:

1. Looking at your data

ds9:

files will automatically open in ds9 if you click on them in Finder

data is stored in 'obs_data'

this folder is conveniently found in any Finder window on the top left

2. Changing the PA

The default PA is East-West (90 degrees, which should be included in your target list regardless). To use different PAs, you have to input it into the OT (and should already be in there for most programs). It is best do this before you slew to the target. If you need to change it even after you have slewed to the target, then change the value in the OT, synch, and add to the OT queue. The SOS will then have to slew again to the target. It will be slow to rotate.

4. Guiding

A. Facility Guiding

For almost all observations, we expect to use the facility guiding at Gemini. 

B. Non-facility Guiding

See other IGRINS wiki pages, as this should be highly unusual.

5. Using the ABBA scripts

There are two available scripts to observe an ABBA quad, which will gain the observer time for preparing for their next source, etc.

Simplest (abba_simple_no_guide.script) - no guiding and relying on the tracking, best option to use for relatively bright sources

Simple with guiding (abba_simple.script)  - best option when you also need to use guiding, and is compatible with both on-slit and off-slit guiding.

To use these scripts, in the DTP:

If you decide you need to cancel (change the name of the file or exposure time or something), it will not run again unless you unclick the "Use Script File" and then click it again.

6. Observing at the Parallactic Angle

This is an advanced mode. 

To observe at the fixed parallactic angle:

If you want to double check that you are indeed at the parallactic angle, you can move your target onto A/B then have your TO input a CE offset. Take another SVC image and verify that the target moved in the slit direction.

You are free to observe as usual, nodding in ABs etc.

7. Saving data

We are working on a best method for distributing data to observers. Do not concern yourself with this.

All IGRINS FITS files are 16.8MB. The average file size of a 1 night run is ~32GB, but this varies depending on your choice to automatically save all the SVC files. The spectroscopic files require ~5GB of disk space each night.

Eventually, the data will be available in the Gemini archive.

End of Night:

1. Close the IG Proxy, then the DSC screen sharing.

2. Close the QLP, SVC, DPT, HPK, and the IGRINS Hub windows (and anything else that is opened other than terminal windows).

3. If it is the morning of Mon-Fri (and not a holiday), then please restart the ICC as a preventative measure. Wait a few minutes, and re-open the screen sharing.

if it does not restart, contact Greg, Hwi, and ask the Gemini observer to leave a note that you need someone to turn

otherwise, perform the following:

4. Close the ICC screen sharing. You are done!